|
|
 |
Walkthrough Page 4
|
|
 |
 |
Add a few drops of Impact Air 45® autoclavable lubricant into the turbine chamber and insert the turbine. Use the bur to guide the spindle into the center of the hole. As the turbine begins to seat, apply firm pressure with a wobbling motion to seat the front O-ring. The end of the rear bearing should be flush with the open end of the turbine chamber. |
 |
 |
 |
Place the headcap on the back of the turbine cartridge so that it rests against the head. Using your fingers, turn the headcap clockwise with slight inward pressure until threads engage. Check to see that the cap is threading straight into the head and continue until finger tight. Continue to tighten the cap with a cap wrench while ensuring that the turbine spins freely.
Be sure to tighten headcap securely so that it does not loosen during operation.
If the turbine does not spin when cap is tight, remove turbine and re-inspect O-rings and washers. If problem persists- contact Palisades Dental for technical assistance. |
 |
 |
 |
|
|