Frequently Asked Questions

Q: How do I lubricate my IMPACT AIR 45® handpiece?
A:  For the standard handpiece (2 hole, 4-hole fiberoptic),
      remove handpiece from hose coupling and insert
      2 drops of lubricant into hole of the large silver "drive air"
      tube. For the Quick Disconnect handpiece, inject 2 drops
      of lubricant in the "air drive hole" of the coupler and
      connect to Quick Disconnect handpiece.

Q: What kind of lubricant should be used?
A:  IMPACT AIR 45® autoclavable lubricant is
     recommended. However, any synthetic-based
     lubricant is acceptable.

Q: How many washers should be on the turbine?
A: Two. One thick (.015), and one thin (.005). These Teflon®
     washers aid in preventing air from escaping into the
     surgical field.

Q: What air pressure (in psi) should be used to run the
     IMPACT AIR 45®?

A: 32 to 40 psi is recommended for optimum performance.

Q: Can a chem-clave be used for sterilization?
A:  Yes. However, it is not recommended. It will sterilize the
     handpiece, but the chemicals will shorten the life of the
     turbine.

Q:  I can pull my bur out immediately after installing it.
      Does this mean that the chuck is faulty?

A:  No. When inserting a bur, slide the bur in as far as it will
     go (about halfway), then press the button firmly and slide
     the bur in the rest of the way. Press the bur against a
     hard, flat surface. This aids in setting chuck grip into the
     bur. Test the bur for security by attempting to pull it out
     with your fingers. If the bur slips without considerable
     force, the chuck should be replaced.

Q:  Can I use nitrogen to operate my IMPACT AIR 45®?
A:  Yes. See Diagram below.

Q:  What if I don't have a headcap wrench?
A:  Headcap wrenches are enclosed with new handpieces,
     replacement headcaps, and conversion kits. If you no
     longer have your headcap wrench, just contact us and we
     will send you a wrench free of charge.

Q:  Which bur comes with the handpiece?
A:  A #1702 friction grip surgical bur.

Q:  Can I have my handpiece repaired by Palisades
      Dental?

A:  Yes. Repairs can be sent directly to Palisades Dental:

       Palisades Dental
       111 Cedar Lane
       Englewood, NJ 07631
       Attn: Repairs

Please see the Repair Page for more information.